|
Problem Descriptions/Solutions
We present here the possible causes and solutions for some of the common
problems that may occur when operating a CM200FEG TEM. The information provided
are collected and compiled from standard sources, such as Philips, as well
as the expertise of experienced users of this lab. If you come to a problem,
please check the list in the frame to the left. Click appropriate button and
the descriptions and solutions will pop up in this frame. However, one may
experience problems that are not listed here. Should you have solved such
problems successfully, we would appreciate your sharing with us the solutions.
Program Hang-up.
Due to a variety of circumstances, the control program may
experience a hang-up. If this occurs, proceed as follows:
1. Press STANBY (may not do anything), then RESTART (recessed button
on top left of panel). Wait for the full microscope start up. When prompted
(at least 20 minutes for vacuum startup), restart the FEG: warm start ok
if less than 2 hrs since hitting RESTART).
2. If startup fails during internal checking of 5 program blocks, try
again. If startup fails again, then download the CM Operating system
as follows: Open program "CMDownload" on PC. Choose file cm125.exe. After program
blocks fail, EM prompts "waiting for PC connection". At this time,
initiate download with CMDownload program. After completion, sometimes
EM restarts automatically, other times you must hit RESTART button. Maybe
you even have to hit RESTART another time. Keep trying until successful
or you suspect a hardware failure. 3. Prior to use, you must reset various
parameters from default values (kV to 200, serial port to enabled, film
number). 4. Check the overall instrument alignment. If the instrument performs
as expected (unlikely), continue with normal operation. If the alignment
appears to be lost, restore using 2 PC programs: FEG Registers (for gun
alignments) and CM Monitor (for column alignments). Use recent files for
restoration (this behooves you to also store these alignments at periodic
intervals).
No High Tension. If everything else is normal, check the 24 V power supply
inside the Power Supply cabinet. In order to test this, open the left-hand
door with the key provided and check the two LEDs on the two lower rows of
the power supplies. THese should be lit. If one of them is not lit, loosen
then four screws approx. 2 cm and pull the unit out. Leave it disconnected
in this position for 30 seconds. After this period, push the unit back into
the rack, return to the operating panel and re-active the H.T. button.
No Emission. This could be due to one of the following:
1. If H.T. conditioning is selected on the first PARAMETERS page, it is
not possible to heat up the filament.
2. After an extremely strong H.T. flashover, the 24 V filament current
power supply in the power suplly cabinet may have switched off. If this is
the case, continue as described in No High Tension, or if the control program
is hang-up, proceed with first as described in Program Hang-up.
3. Filament has broken down. Change filament.
Spot Size Limited. When there is no specimen holder in the microscope, the
spot size will be set to 5 and a spot size indication of XXX will appear on
the microcontroller screen. As soon as a holder is introduced into the goniometer
airlock, the spot size can be controlled again.
No Lens Currents (I). If there are no lens currents measurable when DISPLAY
CURRENTS is selected on the PARAMETERS page, the column lifting circuit which
automatically switches off the lens currents and HT may be active. Unauthorised
attempts to operate the column lifting mechanism can activate the microswitch
which controls the safety circuitry. This point should be checked if the lens
currents connot be obtained when starting to use the microscope. Otherwise,
check first water flow and if necessary reset it to the corrent value.
No Lens Currents (II). This occurred on October 24, 1996. All lens currents
shut off and the cooling water floating sensors dropped off. Action: Remount
the sensors and set their positions. Check the following settings:
1. The water floating level should stay in the range between two the red
markers.
2. The floating sensor should be mounted near the bottom of the floating
gauge.
3. All four floating sensors are connected in series to the safety circuit
with 24V voltage. Checking the voltage between each of the sensor connctors
and ground will identify the malfunctioning sensor.
Control Panel Dead. If this condition occurs, first check the main power
swithc then the fuses behind the right hand door of the power supply cabinet.
Replace any blown fuses and reset the main power switch to on. If all LEDs
on the operating panels are unlit, the ROOM DIM knob may be in the off position.
Vacuum Leak IGP. If IGP stays abnormally high after insertion of any holder
or without a holder, there is probably a small leak in the goinionmeter. If
holder inserted, pull out to intermediate position and reinsert. If no holder,
put one in (and take out again if desired). If cold trap is chilled, then
IGP should recover relatively quickly (5 min). If cold trap not chilled, recovery
would take much longer (best to chill it at this point to make sure everything
ok).
Vacuum Leak P3. It occured on 8/13/96 when taking an exposure. The camera
chamber vacuum reading P3 dramatically increased during a plate transport
(from 52 to 72). The P3 reached 99 after three exposures and the HT & filament
was shut off. The Philips service engineer Joe Maniscalco fixed the problem
with following procedure:
1. Tested plate transport with and w/o plates; same.
2. Dismounted plate transport mechanism, finding no damage on O-ring or
piston.
3. Cleaned and lubbed the O-ring and seats.
4. Reassembled and tested. P3 = 55 with no change during or after plate
transport.
Camera Jam. When taking an exposure, the film plate may be jammed, such as
what happened on 9/30/96, 10/7/96, and several other times. Do the following
to correct the problem.
1. Vent the camera as you would do when change the camera. Check Basic
Procedures regarding how to change the camera.
2. Remove HT cable holder from the supporter connected on the back column
cover.
3. Remove the back column cover.
4. Open the left viewing chamber window glass (Caution: always wear gloves
when work inside the chamber!).
5. Switch Y20.0 and Y20.1 connectors (can be reached from the back of the
microscope right behind the column above the camera level) to release the
force on the plates.
6. Remove the top plate. Serveral attemptes with some manipulations.
7. Reengage the plate transfer stage with the piston arm.
8. Retract the bottom plate back to the camera by switching Y20.0 and Y20.1
forth and back.
9. Switch the Y20.0 and Y20.1 connectors to the original position and put
back the viewing window glass.
Comments: If the jamming is not very severe and only one plate involved,
it is not necessary to open the viewing window. Simply switching the Y20.0
and Y20.1 connectors may clear the jam.
|