Power-up Procedure

1. Let system be at 20kV 10mA until IG reading is substantially below 100mV.

2. Keep TUBE VOLTAGE at 20kV, increase (push UP button) TUBE CURRENT by 10mA once a time until reaching 40mA, and each time wait until IG reading drops back to below 100mV before going to next step.

3. Do the same to gradually go to 30kV 60mA, 40kV 80mA, and finally 50kV 100mA. Again, Make sure IG reading drops below 100mV each time. If IG reading fluctuates dramatically when powering-up, go back to the previous setting (use DOWN button).
See can't power up if failed.


click on image to enlarge view


When system reaches this reading, you can start beam alignment. However, please be aware of the following:

1. The time period for powering-up can vary. Usually, if vacuum was broken (such as after filament change), it will take much longer, sometimes even overnight at intermediate steps.

2. NEVER try to power-up fully when IG reading is greater than 125mV. Normally, the reading stays between 50 and 90mV when the system has stabilized.

3. If for a long period (such as overnight or longer) no user plans to collect data, set the system at 50kV 50mA. This substantially increases filament life time.


click on image to enlarge view


Common Procedures
X-ray generation
Power-up
Change Belt
Polish Anode
Change Water Seal
Change Filament
Change Megnetic Cassette Seal
Pump Release Vacuum
Common Problems
Can't Get X-Ray
Unable to Power-up
Belt Noise
Dirty Anode
Leaking Water
Fluctuating Current
Anode Noise
Bad Vacuum
Machine Down on Anode Rotation
Resources
print this page print this page
e-mail this page e-mail this page
seminars seminars