The whole goniometer assembly is composed of the
base, the goniometer itself and the head. The goniometer is mounted
on the base, which can be shifted horizontally in a direction perpendicular
to the x-ray beam. The lower section of the goniometer is firmly
mounted on the base and thus can only move together with the base.
The upper section of the goniometer can be rotated with respect
to the lower section. The rotation can only be done with the R-AXIS
software. The goniometer head (there are different types) can be
taken off from the goniometer. The head provides additional flexibility
for aligning a sample, which is mounted on the very top of the head.
The top of the head can be translated horizontally. Furtheremore,
the top can also be moved up-and-down.
The most critical thing about the goniometer is that
its rotation axis must intercept perfectly with the x-ray beam and
this is probably the only thing that requires frenquent check.
Check out goniometer alignment for detailed alignment
procedure. |
|
 |
|
|